WANG Zhong-dong, WANG Yu-tian. Principle and design of a test system for pesticide fluorescence lifetime[J]. Journal of Applied Optics, 2005, 26(6): 28-31.
Citation: WANG Zhong-dong, WANG Yu-tian. Principle and design of a test system for pesticide fluorescence lifetime[J]. Journal of Applied Optics, 2005, 26(6): 28-31.

Principle and design of a test system for pesticide fluorescence lifetime

More Information
  • The generation of fluorescence,the mechanism of fluorescence lifetime and the basic measuring principle for fluorescence lifetime are introduced. A test system for fluorescence lifetimes of pesticides is designed by means of the direct record method(the Photon Counting Technique). According to the natures of the samples under test, a corresponding pulsed light source, optical elements, a semiconducting detector, etc. were chosen. All the working parameters such as the pulsed period of the excitation light source were reasonably optimized, a simple modular design was employed, the fluorescence lifetime of carbaryl solution was measured without the interference of the excitation light, and the fluorescence decay curves and the lifetime of 300~400ps (FWHM) were obtained as the concentration of the carbaryl solution is 500μg/L. The results indicate that the system has the virtues of the simplicity in structure and convenience to operate, and can measure lifetimes as short as the order of 100ps.It is suitable for the quantitative measurement fluorescence lifetime of pesticides which can emit fluorescence.
  • Related Articles

    [1]TIAN Ruirui, YANG Peiyu, HU Bojin, LI Yiyang, XU Can. Second-order nonlinear optical response based on codoped C32 multipole moments modulation[J]. Journal of Applied Optics, 2023, 44(4): 726-735. DOI: 10.5768/JAO202344.0401004
    [2]ZHANG Xuecheng, JIN Shangzhong, ZHAO Tianqi, ZHANG Fei, CHEN Yi. Application of aerosol fluorescence spectrum recognition based on deep learning[J]. Journal of Applied Optics, 2022, 43(3): 466-471. DOI: 10.5768/JAO202243.0303001
    [3]Li Jie, Zhang Xiaoqing, Jia Yudong. Design of spectrum measurement system based on Mach-Zehnder optical fiber interference[J]. Journal of Applied Optics, 2017, 38(4): 679-684. DOI: 10.5768/JAO201738.0408002
    [4]Chen Yu, Zhang Zhengping, Wang Jing, Chen Xiaopeng, Zhu Ziyang. Fluorescence spectrum denoising of quantum dots based on improved wavelet threshold[J]. Journal of Applied Optics, 2016, 37(3): 397-401. DOI: 10.5768/JAO201637.0302006
    [5]XIAO Shao-rong, WANG Ya-ji. Inversion of NO2 concentration based on solar spectrum analysis[J]. Journal of Applied Optics, 2011, 32(3): 472-476.
    [6]CHEN Jiao, JIAO Ming-yin, CHANG Wei-jun, KANG Wen-li. Optical design of microscopic imaging system for ultraviolet-visiblewide spectrum[J]. Journal of Applied Optics, 2011, 32(2): 195-199.
    [7]WANG Mei-qin, WANG Zhong-hou, BAI Jia-guang. Removing secondary spectrum in wide spectrum optical system[J]. Journal of Applied Optics, 2010, 31(3): 360-364.
    [8]HUANG Ying, XIE Yan-hong, YI Xin-jian. Banknote recognition technology based on fluorescence spectrum analysis method[J]. Journal of Applied Optics, 2008, 29(4): 629-632.
    [9]WANG Yu-tian, WANG Zhong-dong. Study on Fluorescence Spectrometer for Monitoring Pesticide Residues in Vegetables[J]. Journal of Applied Optics, 2005, 26(5): 10-012.
    [10]WANG Shu-tao, WANG Yu-tian, CHE Ren-sheng, WANG Xiang, JIN Hai-long. Research on the System Based on Wavelet Transform for Measuring Fluorescence Spectrum of Chlorophyll-a/b[J]. Journal of Applied Optics, 2005, 26(1): 49-52.

Catalog

    Article views (2661) PDF downloads (925) Cited by()

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return