WANG Yu-tian, WANG Zhong-dong. Study on Fluorescence Spectrometer for Monitoring Pesticide Residues in Vegetables[J]. Journal of Applied Optics, 2005, 26(5): 10-012.
Citation: WANG Yu-tian, WANG Zhong-dong. Study on Fluorescence Spectrometer for Monitoring Pesticide Residues in Vegetables[J]. Journal of Applied Optics, 2005, 26(5): 10-012.

Study on Fluorescence Spectrometer for Monitoring Pesticide Residues in Vegetables

  • A fluorescence spectrometer which can measure pesticide concentration is designed by using a xenon lamp as excitation lightsource,employing optical fibers to transmit and detect fluorescence and combining with CCD spectrumdetermination technology based on fluorescent characteristics of carbaryl that can emit fluorescence as it is excited by UV rays.Moreover,the rapid measurement of carbaryl residues in cucumbers is implemented with a steadystate fluorescent spectrograph and the new system respectively. The results indicate that the linear concentration range is 0.0~120.0μg/L for carbaryl, the detection limit (LOD) is 5×10-7μg/L,the linear correlation coefficient r is 0.9991 as the excitation wavelength is 319nm and the fluorescence emission wavelength is 647nm. The data prove up to the hilt that the instrument meets the demand of fluorescence detection.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return