CHENG Hong-chang, SHENG Liang, SHI Feng, WANG Fen-fen, FENG Liu. Spectral response measurement of UV image intensifier[J]. Journal of Applied Optics, 2007, 28(3): 305-308.
Citation: CHENG Hong-chang, SHENG Liang, SHI Feng, WANG Fen-fen, FENG Liu. Spectral response measurement of UV image intensifier[J]. Journal of Applied Optics, 2007, 28(3): 305-308.

Spectral response measurement of UV image intensifier

More Information
  • Corresponding author:

    CHENG Hong-chang

  • An ultraviolet spectral response measurement device was designed for measuring the UV image intensifiers fabricated in this laboratory. This system consists of a light source chamber, a grating spectrometer, a DC voltage regulator and a piampere current meter. The radiant power corresponding to every wavelength of the light source was directly read out with the software of the grating spectrometer, the photocurrent generated from a photocathode was directly measured by the piampere current meter when the photocathode was radiated by UV radiation sources, the ratio of the two measurements was then calculated and processed by Microsoft Excel 2003, and relative spectral response curve was obtained. The curve indicates that the spectral response of UV image intensifiers is from 200nm to 340nm and the peak response is near 270nm, which indicates that the UV image intensifier is solar blind. Measurement results show that the uncertainty of the device is less than 10%.
  • Related Articles

    [1]LI Yan, FAN Jihong, YU Bing, YUAN Lin'guang, SUN Yu'nan, QIN Yan, MA Li. Spectral responsivity measurement technology for UV detector based on cryogenic radiometer[J]. Journal of Applied Optics, 2022, 43(2): 311-316. DOI: 10.5768/JAO202243.0204001
    [2]WANG Shengyun, XIE Qi, SHI Jifang, SUN Yunan, YU Bing, KANG Dengkui. Measurement system of radiation sensitivity for UV image intensifier[J]. Journal of Applied Optics, 2020, 41(3): 548-552. DOI: 10.5768/JAO202041.0303001
    [3]CAO Xi-bin, ZHAO Bao-sheng, SAI Xiao-feng, WEI Yong-lin, ZHU Xiang-ping, ZHOU Wei, WANG Jun-feng, DONG Gai-yun. Ultraviolet image intensifier with response at 120 nm~200 nm[J]. Journal of Applied Optics, 2008, 29(5): 731-734.
    [4]XIANG Shi-ming. Theoretical limit for photocathode sensitivityof image intensifier[J]. Journal of Applied Optics, 2008, 29(1): 48-51.
    [5]TAO Kun-yu, LI Fu-wei, FU Sen, ZHOU Yan-ping. Error analysis and comparison of spectral responsivity calibration methods for HgCdTe-IRFPA imaging detector[J]. Journal of Applied Optics, 2007, 28(5): 587-592.
    [6]WANG Ji, ZHENG Xiao-bing, ZHANG Lei, LIN Zhi-qiang. Measurement of spectral responsivity of an infrared detector[J]. Journal of Applied Optics, 2007, 28(3): 313-316.
    [7]FAN Ji-hong, HOU Xi-qi, YANG Zhao-jin, YIN Tao, QIN Yan, LIU Jian-ping. Measurement technology for spectral responsivity of infrared detector[J]. Journal of Applied Optics, 2006, 27(5): 460-462.
    [8]FAN Ji-hong, YANG Zhao-jin, QIN Yan. Research of test technology for absolute spectral responsivity of HgCdTe detector[J]. Journal of Applied Optics, 2006, 27(supp): 79-82.
    [9]SHI Ji-fang, HOU Xi-qi, FAN Ji-hong, SUN Yu-nan. Study of spectral responsivity testing technologyfor photocathodes of image intensifiers[J]. Journal of Applied Optics, 2006, 27(supp): 68-70.
    [10]ZHAN Chun-lian, LI Yan-mei, LIU Jian-ping, LI Zheng-qi. The Unitormity and Linearity of Infrared Spectral Response of Decector[J]. Journal of Applied Optics, 2004, 25(6): 34-37.

Catalog

    Article views (3408) PDF downloads (1151) Cited by()

    /