WU Xing-lin, QIU Ya-feng, QIAN Yun-sheng, LIU Zhao-lu, CHENG Hong-chang. Relationship between voltage of MCP and signal-to-noise ratio of UV image intensifier[J]. Journal of Applied Optics, 2013, 34(3): 494-497.
Citation: WU Xing-lin, QIU Ya-feng, QIAN Yun-sheng, LIU Zhao-lu, CHENG Hong-chang. Relationship between voltage of MCP and signal-to-noise ratio of UV image intensifier[J]. Journal of Applied Optics, 2013, 34(3): 494-497.

Relationship between voltage of MCP and signal-to-noise ratio of UV image intensifier

  • Ultraviolet image intensifier has a very broad application prospects in many fields, such as space optical detection, environmental monitoring, strategic defense, and so on. Signal-to-noise ratio measurement system of ultraviolet image intensifier was designed by reference to the test principle of signal-tonoise ratio for UV image intensifier, and its test conditions were provided. The relationship between voltage of MCP and signal-to-noise ratio of UV image intensifier was deeply studied, and its curve is measured by using the development of its own signal-to-noise ratio tester for UV image intensifier. The result shows that in the region of less than 850 V, the output of the signal-to-noise ratio of the ultraviolet image intensifier is increased with the increase of voltage; when the voltage is more than 850 V, the output of the signal-to-noise ratio of the ultraviolet image intensifier is essentially unchanged.