WU Zhi-feng, DAI Cai-hong, YU Jia-lin, HUANG Bo, OUYANG Hui-quan. Performance investigation of blackbody BB3 500 M and temperature measurement[J]. Journal of Applied Optics, 2012, 33(5): 926-930.
Citation: WU Zhi-feng, DAI Cai-hong, YU Jia-lin, HUANG Bo, OUYANG Hui-quan. Performance investigation of blackbody BB3 500 M and temperature measurement[J]. Journal of Applied Optics, 2012, 33(5): 926-930.

Performance investigation of blackbody BB3 500 M and temperature measurement

More Information
  • High temperature blackbody BB 3 500 M was used as the primary standard of spectral irradiance at national institute of metrology (NIM). The blackbody BB 3 500 M cavity radiator was consisted of a series of pyrolytic-graphite (PG) rings, capable of operating at temperature up to 3 500 K. Feedback control mode was used to stabilize the temperature of the blackbody. The stability of the radiator and the blackbody uniformity were measured. In transferring the spectral irradiance from the blackbody to the working standard lamp, the temperature of the blackbody may change. In the experiment, temperature drift within an hour was tested. When the blackbody was heated up to around 3 016 K, the temperature uncertainty was less than 0.3 K. In the radiation path, an aperture was used to screen the radiation from the cylindrical part of the radiation cavity. However, the temperature was not the same all over the effective area of the graphite bottom. The uniformity of the bottom of BB 3 500 M was found to be within 0.2 K. Results showed that BB 3 500 M was suitable for the primary radiation source. In the temperature measurement, temperature below 2 473 K could be traceable to the heat division of NIM. Then the temperature could be extended through the comparison of the spectral radiance at fixed wavelength. In the experiment, the temperature extend was first investigated in low temperature region no higher than 2 473 to verify its applicability. Results in high temperature region showed good consistence that the temperature divergence was smaller than 0.5 K at 345 nm~360 nm wavelength.
  • [1]WOOLLIAMS E R, FOX N P, COX M G, et al. The CCPR K1-a key comparison of spectral irradiance from 250 nm to 2 500 nm: measurements analysis and results[J]. Metrology, 2006, 43:S98-S104.
    [2]KLEIN R, TAUBERTt D, THORNAGEL R, et al. Radiometric comparison of the primary synchrotron radiation source standard metrology light source with calibrated filter radiometers in the visible and NIR spectral range[J]. Metrology, 2009, 46:359-366.
    [3]SHAW P S, ARP U, SAUNDERS R D, et al. Synchrotron radiation-based irradiance calibration from 200 to 400 nm at the synchrotron ultraviolet radiation facility III[J]. Applied Optics, 2007, 46:25-35.
    [4]RICHTER  M, HOLLANDT J, KROTH U, et al. Source and detector calibration in the UV and VUV at bessy II[J]. Metrology, 2003, 40:S107-S110.
    [5]ZWINKELS J C, IKONEN E, FOX N P, et al. Photometry, radiometry and ‘the candela: evolution in the classical and quantum world[J]. Metrology, 2010, 47 :R15-R32.
     [6]WOOLLIAMS E R, HARRISON N J, FOX N P. Preliminary results of the investigation of a 3 500 K black body[J]. Metrology, 2000, 37:501-504.
    [7]SPERFELD P, METZDORF J, HARRISON N J, et al. Investigation of high-temperature black body BB3200[J]. Metrology, 1998, 35: 419-422.
    [8]KLEIN R, GOTTWALD A, BRANDT G, et al. Comparison of the primary source standard ‘metrology light source- to a primary detector standard[J]. Metrology, 2011, 48:219-225.
    [9]占春连,刘建平,李正琪,等. 基于高温黑体的辐射亮度的研究[J]. 应用光学,2006, 27(增刊):71-75.
    ZHAN Chun-lian, LIU Jian-ping, LI Zheng-qi, et al. Research on measurement of spectral radiance luminance based on high temperature blackbody[J]. Journal of Applied Optics, 2006, 27(sup): 71-75. (in Chinese with on English abstract)
    [10]代彩红,于家琳. 用单色亮度比较法确定高温黑体的温度[J]. 光学技术,2002, 28(03):250-252.
    DAI Cai-hong, YU Jia-lin. Determination high temperature of blackbody using monochromator radiance ratio method[J]. Optical Technique, 2002, 28(03):250-252. (in Chinese with on English abstract)
  • Related Articles

    [1]WANG Gang, WAN Xun, CUI Zhichao, XIE Liangping. Control scheme of high working temperature based on dynamic temperature control in FOG[J]. Journal of Applied Optics, 2023, 44(5): 1153-1156. DOI: 10.5768/JAO202344.0508004
    [2]YUAN Liang, YUAN Lin'guang, DONG Zaitian, LI Yan, FAN Jihong, LU Fei, ZHAO Juncheng, ZHANG Deng, YOU Yue. Measurement of normal spectral emissivity of materials at high temperature[J]. Journal of Applied Optics, 2023, 44(3): 580-585. DOI: 10.5768/JAO202344.0303002
    [3]HU Tieli, WANG Honghong, LI Siwei, CAO Feng, HU Xinyi, FAN Zhe, YANG Yuxin, GUO Jian, YOU Yue, YANG Ke, LI Hui, YU Yang. Research on temperature control and self-tuning for 30℃~420℃ blackbody[J]. Journal of Applied Optics, 2023, 44(2): 392-397. DOI: 10.5768/JAO202344.0203005
    [4]ZHANG Jing, LI Yongqian. Temperature sensing characteristics based on coreless- few mode-coreless optical fiber structure[J]. Journal of Applied Optics, 2022, 43(1): 167-170. DOI: 10.5768/JAO202243.0108003
    [5]HE Shufang, WANG Yanfei, DAI Caihong, LIU Jinyuan, FENG Guojin. Research of temperature uniformity and size-of-source effect in infrared spectral radiance measurement[J]. Journal of Applied Optics, 2020, 41(4): 737-742. DOI: 10.5768/JAO202041.0406003
    [6]Yu Dongyu, Zhang Kuijia, Wu Yao, Zhu Yinsheng, Wu Pei, Yin Wanhong, Li Siwei, Yue Wenlong. Temperature measurement of blackbody using filter radiometer[J]. Journal of Applied Optics, 2018, 39(6): 862-866. DOI: 10.5768/JAO201839.0603002
    [7]ZHAI Yang, ZHU Ri-hong, SHEN Hua, GU Jin-liang. Multi-spectral radiation based transient pyrometer for high-speed dynamic system measurement[J]. Journal of Applied Optics, 2011, 32(4): 698-704.
    [8]LI Hao, ZHANG Yan-ge. Atmospheric temperature measurement with Fabry-Perot airglow interferometer imaging system[J]. Journal of Applied Optics, 2008, 29(6): 921-925.
    [9]ZHAN Chun-lian, LIU Jian-ping, LI Zheng-qi, LU Fei, CHEN Chao. Research on measurement of spectral radiance luminance base on hightemperature blackbody[J]. Journal of Applied Optics, 2006, 27(supp): 71-75.
    [10]LIU Jian-ping, ZHANG Hui, ZHAN Chun-lian, LI Zheng-qi. Mathematical model and method for measurement of spectral irradiance[J]. Journal of Applied Optics, 2005, 26(6): 70-073.

Catalog

    Article views (2746) PDF downloads (497) Cited by()

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return