MRTD in multispectral infrared imaging system based on AOTF
-
-
Abstract
In order to achieve the spectral minimum resolvable temperature difference (MRTD) testing of infrared multispectral imaging and obtain more information of target and background, an acoustooptic tunable filter (AOTF)based multispectral static performance testing platform was designed and developed. At 3 m~5 m waveband,through adopting the designed optical testing platform,the subdivision of band was carried out in middlewave infrared region,and the MRTD tests were finished in targets with different spatial frequencies of 2.546 lp/mm,1.273 lp/mm,0.849 lp/mm and 0.632 lp/mm,respectively. The experimental result shows that, the test results of spectral MRTD correspond with the theoretic analysis,verifying the reliability of the optical testing platform and the feasibility of the testing method.
-
-