Abstract:
Image luminance meter can effectively detect luminance, chroma and spatial light distribution of LED display light source. However, when testing monochrome LED, due to its narrow spectral bandwidth, traditional image luminance meter calibrated with standard A light source has spectral mismatch problem, which leads to large test error. In order to solve this problem, this paper studied the test method of image luminance meter based on output spectra of LED light sources. Firstly, the influence of different LED colors on test results was analyzed, and an improved convolutional neural network was constructed based on Alexnet model to realize light source color recognition; on this basis, the mathematical model of luminance and chroma was established to correct and calculate luminance and chroma of light source. Secondly, the mathematical relationship between grayscale signal output by image luminance meter and exposure time of camera was studied. Results indicate that the two are linearly related, which provides an important basis for test calibration. Different from traditional standard A light source calibration method, this method uses multiple spectra output by LED light sources for testing, and finally controled test error of LED luminance and chroma within 5%, which provides an effective solution for improving measurement accuracy of image luminance meters.